Operating the calibrator

2.1

Before use

Warning
The calibrator must not be used for any purposes other
than those described in this manual.
The calibrator is designed for interior use only and
should not be used in risk-prone areas, where vapour
or gas leaks, etc. may constitute an explosives hazard.
The connections used to test thermostats (Fig. 1 pos.
10 and 11) must NEVER be connected to voltages
exceeding 50V. The thermostats must not be connected
to any other source of voltage during the test.
NEVER use heat transfer fluids such as silicone, oil,
paste, etc.
The calibrator must be kept clear within an area of 20
cm on all sides and 1 metre above the calibrator.
Use only insulation plugs supplied by Ametek Denmark
A/S. Do not try to modify the insulation plugs (CTC-
1200 A only).
Caution – Hot surface
This symbol is engraved in the grid plate.
Do not touch the grid plate, the well or the insertion
tube as the calibrator is heating up – they may be very
hot.
Do not touch the handle of the calibrator during use – it
may be very hot.