What If; Error Rectification - Ivoclar Vivadent Bluephase Style 20i Instrucciones De Uso

Ocultar thumbs Ver también para Bluephase Style 20i:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 74

6. What if ...?

Indicator
Causes
The device is overheated.
All LED's
orange
Electronic component of the handpiece
All LED's
red
is defective.
– Power pack not connected or
The charging
defective
base is not
– Battery fully charged
illuminated
during charging
Battery empty
No indicator
activity of the
device with the
Battery contacts dirty
battery in place
18

Error rectification

Allow the device to cool down and try again after a
certain time. If the error persists, please contact your
dealer or your local Service Centre.
Remove and reinsert the battery. If the error persits,
please contact your dealer or your local Service
Centre.
Check if the power pack is correctly positioned in
the charging base or if the power pack is connected
to the power supply by means of the power cord
(LED on the power pack lights up in green if it works
correctly).
Place the device in the charging base and charge
for at least 2 hours.
Remove battery and clean the battery contacts.

Publicidad

Tabla de contenido
loading

Tabla de contenido