Disinfection
Clean all parts of the mask before disinfecting it as described below.
PASTEURIZATION
Temperature:
70°C
Exposure time:
30 minutes
Result:
High level disinfection
Maximum allowable
cleaning cycles:
10
CIDEX
®
Exposure time:
45 minutes
Rinse:
Per manufacturer's
instructions
Result:
High level disinfection
Maximum allowable
cleaning cycles:
10
CIDEX O. P. A.
®
Exposure time:
12 minutes
Rinse:
Per manufacturer's
instructions
Result:
High level disinfection
Maximum allowable
cleaning cycles:
10
STERRAD
System 100
®
Exposure time:
Machine controlled
Result:
Sterilization
Maximum allowable
cleaning cycles:
10
If the swivel does not move freely after disinfection, replace the sealing ring (part no.
452072).
Verify function of the flapper valve after processing. If the flapper does not move freely,
replace the mask.
Inspect the mask after processing. If any componenets are damaged, replace the Total
Face Mask.
If the swivel does not move freely after disinfection, replace the sealing ring (part no.
452072).
Verify function of the flapper valve after processing. If the flapper does not move freely,
replace the mask.
Inspect the mask after processing. If any componenets are damaged, replace the Total
Face Mask.
Slight discoloration of the mask cushion after processing is normal.
If the swivel does not move freely after disinfection, replace the sealing ring (part no.
452072).
Verify function of the flapper valve after processing. If the flapper does not move freely,
replace the mask.
Inspect the mask after processing. If any componenets are damaged, replace the Total
Face Mask.
Slight discoloration of the mask cushion after processing is normal.
If the swivel does not move freely after disinfection, replace the sealing ring (part no.
452072).
Verify function of the flapper valve after processing. If the flapper does not move freely,
replace the mask.
Inspect the mask after processing. If any componenets are damaged, replace the Total
Face Mask.
A slight odor may be present after processing. This is normal.