DeVilbiss Healthcare 515A Serie Guía De Instrucciones página 8

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 10
catalog Number
Operating System
For units equipped with an OSD, the
visible "low oxygen" indicator will
activate at the following level
Storage Conditions
Equipment Class and Type
Approval Body and Safety Standard
EmC Compliance To
*NOTE: The OSD performance at 10°C to 35°C, 95% R.H. through voltage range on the 515ADS/515AKS/515DS verified at 670m.
**NOTE: This corresponds with a sound level of 50 dbA overall average as measured according to the guidelines of the now defunct ANSI Z79.123-1981 standard.
Specifications subject to change without notice.
EN - 8
DEvILbISS 4-LItER AND 5-LItER SERIES
515ADS/515ADZ
515DS/515DZ
Time Cycle / Pressure Swing
515A Series
84% ± 2%
(The audible alarm will alert at
approximately 75%. At less than 60%, the
red "service required" light will activate.)
515 Series
Units with serial numbers less than H20000:
85%± 2% (The audible alarm will alert at
approximately 75%).
Units with serial numbers of H20000 and
greater: 83.5%± 1.5%. (The audible alarm
will alert at approximately 75%. At less than
60%, the red "service required" light will
activate.)
-40°C to 70°C, humidity range of 10% to
100%, including condensation
Class II Equipment Double Insulated;
Type B Applied Part
CSA
CAN/CSA-C22.2 No. 601.1-M90
515A Series - IEC 60601-1-2
515 Series - IEC 601-1-2
515AKS/515AKZ
515KS/515KZ/515NS
Time Cycle / Pressure Swing
515A Series
84% ± 2%
(The audible alarm will alert at
approximately 75%. At less than 60%, the
red "service required" light will activate.)
515 Series
Units with serial numbers less than H20000:
85%± 2% (The audible alarm will alert at
approximately 75%).
Units with serial numbers of H20000 and
greater: 83.5%± 1.5%. (The audible alarm
will alert at approximately 75%. At less than
60%, the red "service required" light will
activate.)
-40°C to 70°C, humidity range of 10% to
100%, including condensation
Class II Equipment Double Insulated;
Type B Applied Part
515A Series
TUV
IEC 60601-1
EN 60601+A1+A2
ISO8359:1996 (515AKZ excluded)
515 Series
TUV
IEC 601-1+A1+A2
EN 60601+A1+A2
ISO8359:1996 (515KZ excluded)
515A Series - IEC 60601-1-2
515 Series - IEC 601-1-2
515UK 4-Liter
Time Cycle / Pressure Swing
Units with serial numbers less than
H20000: 85%± 2% (The audible alarm
will alert at approximately 75%).
Units with serial numbers of H20000
and greater: 83.5%± 1.5%. (The
audible alarm will alert at
approximately 75%. At less than 60%,
the red "service required" light will
activate.)
-40°C to 70°C, humidity range of 10%
to 100%, including condensation
Class II Equipment Double
Insulated;
Type B Applied Part
TUV
IEC 601-1+A1+A2
EN 60601+A1+A2
ISO8359:1996
IEC 601-1-2
SE-515X

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

515 serie

Tabla de contenido