Press and hold the ON/OFF button 1 for 8 seconds until
«CL Pr» is displayed. This clears the Bluetooth
records on the device and reset it to factory settings. Blue-
®
tooth
automatically activates after the reset.
For more detailed information visit www.microlife.com/connect.
9. PC-Link Functions
This device can be used in conjunction with a personal computer
(PC) running the Microlife Blood Pressure Analyzer+ (BPA+) soft-
ware. The memory data can be transferred to the PC by
connecting the monitor via a cable.
If no download-voucher and cable is included download the BPA+
software from www.microlife.com/software and use a Micro-USB
cable.
10. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
Error
Description Potential cause and remedy
«Err 1»
Signal too
The pulse signals on the cuff are too
weak
BK
weak. Re-position the cuff and repeat the
measurement.*
«Err 2»
Error signal During the measurement, error signals were
BT-B
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«Err 3»
Abnormal
An adequate pressure cannot be gener-
BT-C
cuff
ated in the cuff. A leak may have occurred.
pressure
Check that the cuff is correctly connected
and is not too loose. Replace the batteries
if necessary. Repeat the measurement.
«Err 5»
Abnormal
The measuring signals are inaccurate and
result
no result can therefore be displayed. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
«Err 6»
MAM Mode There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
BP B6 Connect
Error
Description Potential cause and remedy
®
pairing
«HI»
Pulse or cuff
pressure too
high
«LO»
Pulse too
low
Problem with
Bluetooth
connection
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
11. Safety, Care, Accuracy Test and Disposal
Safety and protection
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions
described in the «Technical Specifications» section.
Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
The cuffs are sensitive and must be handled with care.
Do not exchange or use any other kind of cuff or cuff connector
for measuring with this device.
Only pump up the cuff once fitted.
Do not use this device close to strong electromagnetic fields such
as mobile telephones or radio installations. Keep a minimum
distance of 3.3 m from such devices when using this device.
Do not use this device if you think it is damaged or notice
anything unusual.
The pressure in the cuff is too high (over
299 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
If any problem occurs with the Bluetooth
®
connection, the Bluetooth
blinks rapidly for approximately 10
seconds. To solve the problem, please
visit www.microlife.com/connect.
®
icon AO
EN