What to do if ...
Problem
Reason
Batteries are flat.
Memory readings can be
called up.
Measurements are not
(flashing)
possible.
e.g.
Error message below 30:
Device may not be correctly
................
positioned on the wrist.
e.g.
Error message 30 and
higher:
................
System error
Measurement or
Right measuring position
inflation does not
was not reached.
start instantly.
Readings from
Blood pressure is a
repeated meas-
fluctuating value.
urements differ
For healthy adults,
considerably.
deviations of about 10 to
20 mmHg are possible.
Cuff is not properly fitted.
Readings were not con-
sistently taken at heart
level.
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
Prior to a meas-
Position sensor is de-
urement, the
activated.
position sensor
icon on the display
does not work.
All manuals and user guides at all-guides.com
Solution
Insert new batteries.
Refit device and repeat measurement
as shown and described under
«Measuring».
Repeat measurement. If problem
persists, have device checked at a
Braun Service Centre.
Repeat measurement as shown and
described under «Measuring» or
deactivate the position sensor.
none
Ensure that the appliance is fitted
as follows:
– on top of the slim side of the forearm
– with the display on top
– with the battery compartment facing
the user's body
– approx. 2 cm away from the thumb
root
For each measurement, proceed
exactly as shown and described
under «Measuring».
While measuring, relax, keep still,
do not move, or speak.
Refer to the chapter «Select mode»
to activate the position sensor.
11