SYMBOLS USED
TYPE BF APPLIED PART: Equipment providing a degree of protection against electric
shock, with isolated applied part. Indicates that this device has conductive contact with
the end user.
This symbol on the unit means "Refer to instructions for use".
Temperature Limitation: indicates the temperature limits to which the medical device
can be safely exposed.
Lot Number: indicates the manufacturer's batch code so that the batch or lot can be
identified.
Humidity Limitation: indicates the humidity limits to which the medical device can be
safely exposed.
Serial Number: indicates the manufacturer's serial number so that a specific medical
device can be identified.
Do not dispose in household waste.
Catalogue Number: indicates the manufacturer's catalogue number so that the device
can be identified.
Atmospheric Pressure: indicates the atmospheric limits to which the medical device
can be safely exposed.
Manufacturer Symbol
Date of Manufacture: indicates the date which the medical device was manufactured.
This is included within the serial number found on the device (usually in the battery
compartment), either as "E/Year/Number" (YY/123456) or "E/Month/Year/Number"
(MM/YY/123456).
This medical device is indicated for home use.
IP22
This medical device is not water resistant and should be protected from liquids.
The first number 2: Protected against access to hazardous parts with a finger, and
the jointed test finger of 12 mm ø, 80 mm length, shall have adequate clearance from
hazardous parts, and protected against solid foreign objects of 12.5 mm ø and greater.
The second number 2: Protected against vertically falling water drops when
enclosure is tilted up to 15˚. Vertically falling drops shall have no harmful effects when
the enclosure is tilted at any angle up to 15˚ on either side of the vertical.
CE Mark
Medical Device
4