HoMedics BPW-1005-EU Manual De Instrucciones página 14

Ocultar thumbs Ver también para BPW-1005-EU:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 15
Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
portable and mobile RF communications equipment and the [BPW-1005]
The [BPW-1005] is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the [BPW-1005] can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the [BPW-1005]
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
All manuals and user guides at all-guides.com
Recommended separation distances between
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2 √
P
0.12
0.38
1.2
3.8
12
14
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2 √
P
0.12
0.38
1.2
3.8
12
d = 2.3 √
P
0.23
0.73
2.3
7.3
23

Publicidad

Tabla de contenido
loading

Tabla de contenido