5 USING THE CONDUCTIVITY METER (continued)
When using the reading locking function :
4
After
appears, press the MEAS button
for 0.5 seconds. The reading locking
function is activated.
the measured value is stabilised. When
the measured value is stable,
ashing and the displayed value is locked
with
and
MEAS
5
Read the displayed value.
6
Press the MEAS button for 0.5 seconds.
The reading locking function is deactivated
and
disappears.
MEAS
Note: If a measurement result is out of the specied measurement range, the
displayed measured value ashes.
5.7 AFTER MEASUREMENT
1
Press the ON/OFF button to switch the meter off.
2
Wash the sensor with tap water and wipe away any residual
water using a clean tissue.
3
Replace the sensor protection cap.
Note: If the meter is to remain unused for a long period of time, use pure water instead
of tap water to wash the sensor.
6 CARE & MAINTENANCE
The Elcometer 138 Conductivity Meter is designed to give many years
reliable service under normal operating conditions.
Prolonged periods of non-use may cause the sensor to dry out.
This can result in malfunction or unstable readings. Pour
moistening solution into the sensor cell and leave for a few
minutes to allow the sensor to become saturated. Wash the
sensor with water prior to use.
If the measuring surface of the sensor is contaminated or if air
bubbles are regularly present in the sample, clean the sensor
using a diluted neutral detergent (diluted 100 times).
e
When using the reading locking function, deactivate the function before starting every measurement.
www.elcometer.com
All manuals and user guides at all-guides.com
ashes until
MEAS
MEAS
illuminated steadily.
e
MEAS
stops
R
en-12