Revert to factory default settings
The continually displayed CAL symbol refers to a previously made
adjustment.
What to do
1.
Bring up the CAL
mode
2.
Deactivate CAL
mode
3.
Revert to factory
setting
Important:
When the Transferpette
offset is performed, which means that the volume is changed
across the entire volume range of the pipette by the same
amount. It is recommend that the adjustment be performed
at 50% of the nominal volume.
Note:
The instrument is permanently adjusted for watery solutions,
but it can also be set for solutions with varying density, vis-
cosity and temperature. The Transferpette
adjusted in every mode, with the exception of the GEL mode.
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
‚on'. ‚CAL' blinks.
Press one of the arrow keys
to deactivate the CAL mode.
The display changes from
‚on' to ‚off'. ‚CAL' continues
to blink.
Press the ENTER key.
The CAL symbol disap-
pears. The instrument has
now been reverted to factory
default setting.
electronic is adjusted, a volume
®
Easy Calibration
Keys to press
Display readout
1x
>3s
1x
electronic can be
®
59