Technical Data - ADE M118600 Instrucciones De Uso

Báscula electrónica para bebés
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 57
Power supply
Mains operation:
Mains voltage:
Frequency:
Supply Voltage:
Current:
Battery operation:
"Weight" Measuring
range
Max. load capacity:
Division:
Accuracy:
"Length" Measuring range
Measuring Range:
Division:
Accuracy:
Ambient conditions
Operating temperature:
Storage and Transport
temperature:
Humidity:
Air pressure:
Device classification
Mode of operation:
Degree of protection:
Battery mode:
AC adapter mode:
Software
Wireless:
Frequency:
Transmission power:
Range:
Housing
Dimensions:
Net weight:
Material:
Protection class:
Service life:
M118600+M118600-01-210305-Rev002-UM-en

9. Technical Data

Use only with mains adapter UES06WOCP-060100SPA
100 – 240 V AC, 0.2A
50/60 Hz
6 V DC
1.0 A (max.)
4 x 1.5 V AA alkaline batteries
(> 5000 measurements / >100 h operating time)
20 kg
5 g
±10 g / ±0,3%
39,8 – 80,2 cm
0,2 cm
±0,4 cm
+10°C to +45°C
-20°C to +60°C
10% - 95% RH
700 hPa - 1060 hPa
The combination of scale and adapter are specified as ME System. Adapter
is considered as non-ME equipment part of ME System.
Continuous operation
Type BF applied part (scale top surface)
Internally powered ME equipment
Class II ME equipment
The software version is displayed when the scale is switched on.
2402 MHz to 2483.5 MHz
+4 dbm
10 metres
610 x 390 x 140 mm
655-1055 x 390 x 140 mm
3,5 kg / 3,7 kg
ABS plastic
IP21; the device is protected against solid foreign objects with a diameter of
≥ 12.5 mm. Protected against vertically falling drops of water or
condensation.
The design provides you with a service lifetime of 8 years.
44,09 lb
0,01 lb
15,625" – 31,625"
0,125" (1/8")
without length measuring system
with length measuring system
without / with length measuring system
Seite 33 von 112

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

M118600-01

Tabla de contenido