14.2 General procedure for cleaning/disinfection

Preparation
▪ Move the bed to its highest position and adjust the backrest slightly.
▪ Disconnect the bed from the power supply.
▪ Lock the brakes and pull up the siderails.
▪ Make sure that the product is free of coarse impurities and dirt particles that could damage
the surface during cleaning/disinfection.
▪ Dose cleaning agents or disinfectants exactly according to the manufacturer's instructions.
▪ Use a soft cloth lightly moistened with all-purpose cleaner/disinfectant solution.
Sequence of components for cleaning/disinfection
▪ Handrails of the head part and footrest
▪ Inside and outside of head part and footrest
▪ Handrails and release buttons of the siderails
▪ Front and rear siderails and telescopic mechanism
▪ Handset including handset hook and holder, if applicable
▪ All accessories attached to the bed
▪ Side panels inside and outside
▪ Top and bottom of the mattress
▪ Patient surface elements, top frame
▪ Scissor frame (undercarriage), brake bracket
Post-preparation
▪ Then wipe the product thoroughly with a soft, slightly damp cloth (water only) to neutralize it.
▪ Make sure that the product is free of cleaning and disinfectant residues.
▪ Dry the product with an absorbent, non-linting cloth.
▪ Make sure that no moisture/wetness has remained on the surfaces or penetrated into gaps.
▪ Reconnect the bed to the power supply.