• When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff tube
should not have any sharp kinks.
9. Specifications
Model no.
BM 20
Measurement
Oscillometric, non­invasive blood pres­
method
sure measurement on the upper arm
Measurement
Cuff pressure 0 – 300 mmHg,
range
systolic 50 – 250 mmHg,
diastolic 40 –180 mmHg,
Pulse 40 –160 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 30 memory spaces
Dimensions
L 110 mm x W 170 mm x H 58 mm
Weight
Approx. 426 g (without batteries)
Cuff size
22 to 32 cm
Permissible operat­
+ 10 °C to + 40 °C, ≤ 85 % relative air
ing conditions
humidity (non­condensing)
Permissible storage
­ 5 °C to + 50 °C, ≤ 85 % relative air
conditions
humidity, 700 – 1060 hPa ambient pres­
sure
All manuals and user guides at all-guides.com
Power supply
Battery life
Accessories
Classification
Technical information is subject to change without notification
to allow for updates.

10. Mains part

Model no.
Input
Output
Supplier
Protection
21
4 x 1,5 V
AA batteries
For approx. 300 measurements, depend­
ing on the blood pressure level and/or
pump pressure
Instruction for use, 4 x 1.5 V AA batteries,
storage pouch
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
8194047/02
100 V – 240 V ~, 50/60 Hz
6 V, 600 mA, only in connection with
beurer blood pressure monitor
Ansmann AG
This device is double isolated and pro­
tected against short circuit and overload
by a primary thermal fuse.
Make sure to take the batteries out of
the compartment before using the mains
part.
Double isolated / equipment class 2