The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification
to allow for updates.
• This unit is in line with European Standard EN 60601-1-2 (In
accordance with CISPR 11, IEC 61000-4-2, IEC 61000-4-3,
IEC 61000-4-6 and IEC 61000-4-8) and is subject to particu-
lar precautions with regard to electromagnetic compatibility
(EMC). Please note that portable and mobile HF communica-
tion systems may interfere with this unit.
• This device is in line with the EU Medical Devices Directive
93/42/EEC, the "Medizinproduktegesetz" (German Medical
Devices Act) and the standards EN 1060-1 (non-invasive
sphygmomanometers, Part 1: General requirements),
EN 1060-3 (non-invasive sphygmomanometers, Part 3:
Supplementary requirements for electro-mechanical blood
pressure measuring systems) and IEC 80601-2-30 (Medical
electrical equipment – Part 2 – 30: Particular requirements
for the safety and essential performance of automated non-
invasive blood pressure monitors).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful
life. If using the device for commercial medical purposes, it
must be regularly tested for accuracy by appropriate means.
Precise instructions for checking accuracy may be requested
from the service address.

11. Mains part

Model no.
Input
Output
Supplier
Protection
Enclosures and
Protective Covers

12. Replacement parts and wearing parts

Replacement parts and wearing parts are available from the
corresponding listed service address under the stated material
number.
29
LXCP12-006060BEH
100 – 240 V, 50 – 60 Hz, 0.5 A max
6 V DC, 600 mA, only in connection with
beurer blood pressure monitor.
Shenzhen Iongxc power supply co., ltd
This device is double insulated and
protected against short circuit and
overload by a primary thermal fuse.
Make sure to take the batteries out
of the compartment before using the
mains part.
Polarity of the the DC voltage
connection
Double insulated / equipment class 2
Equipment enclosed to protect against
contact with live parts, and with parts
which can become live (finger, pin,
hook test).
The operator shall not contact the
patient and the output plug of AC mains
part simultaneously.