Mount
3/32" Hex Wrench
After insertion, tighten
tensioning screw
until desired rotation
resistance is obtained.
Phillips Head
Screwdriver
Tighten Screw with keyed
washer to retain tilter
mechanism.
Monitor Hold Down
Monitor pivot tension adjustment
Monitor tilt tension adjustment
Warning! Monitor
must be latched during
transportation. Failure
to properly secure
monitor may result in
equipment damage or
injury to personnel.
Adjustment Points
www.stryker.com
2 - EN
Operation
Rotational Lock
Adjustments
Should there be a need for adjustment, tighten or loosen
{
any of the 7 adjustment points that are shown.
Take care not to over tighten any of the adjustments.
{
Over tightening will make moving the arm and monitor
{
difficult and may cause premature wear of components.
Periodically check all adjustments to insure proper
{
operation.
WARNING!
Failure to properly adjust arm could result in equipment
damage or injury.
CAUTION!
Do not over tighten tensioning screws. Damage to bushing may
occur.
Periodically ensure that the tensioning screw, Philips head
screw and washer are suitably tightened and flush as directed
in step 5.
Ensure all adjustment points are operating at required tightness
to prevent inadvertent movement.
Upper arm articulation tension adjustments
Horizontal Arm Rotation
Adjustment.
NOTE: Horizontal Arm cover
plate must be removed to
access the adjustment screw.
Rotate lock knob
clockwise to lock,
counterclockwise
to unlock.
LIFT / ROTATE
LEVER/PIVOTER
ARM
COUNTERBALANCE
ADJUSTMENT /
Ajustement
du contrepoids du bras
Upper arm pivot
tension adjustment