HT COMBI5 Serie Guia De Inicio Rapido página 24

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 28
MEASUREMENT OF NETWORK PARAMETERS AND HARMONICS
DC Voltage
Range [V]
15.0  265.0
Values lower 15V are zeroed
AC TRMS Voltage
Range [V]
15.0  459.9
Values lower 15V are zeroed; Max crest factor: 1.5
Frequency
Range [Hz]
47.5  63.0
Allowed voltage range: 5.0  459.9V ; Allowed current range: ≥5mVAC
DC Current with transducer clamp (In1 input – STD clamp)
Range [mV]
-1999.9  -1.0
1.0  1999.9
Values lower to ±1mVDC are zeroed
AC TRMS Current with transducer clamp (In1 input – STD clamp)
Range [mV]
1.0  2999.9
Values lower to 1mVAC are zeroed ; Max crest factor: 3
DC/AC TRMS current with transducer clamp (In1 input – STD clamp)
FS clamp / Output ratio
1A/1V AC
5A/1V AC
10A/1V AC/DC
30A/3V AC
40A/400mV AC/DC
100A/1V AC/DC
200A/1V AC
300A/3V AC
400A/400mV AC/DC
1000A/1V AC/DC
2000A/1V AC
3000A/3V AC
DC Power
FS clamp
 10A
10A  FS  40
40A  FS  100
100A  FS  1000
Active power (@ 230V single phase systems 1Ph, cos=1, f=50/60Hz)
FS clamp
 10A
10A  FS  200
200A  FS  1000
1000A  FS  3000
Reactive power (@ 230V single phase systems 1Ph, cos=0, f=50/60Hz)
FS clamp
 10A
10A  FS  200
200A  FS  1000
1000A  FS  3000
Resolution [V]
0.1V
Frequency [Hz]
50/60Hz ±5%
Resolution [Hz]
0.01
Resolution [mV]
0.1
Frequency [Hz]
50/60Hz ±5%
Measurement range
0.1mA ÷ 999.9mA AC
0.001A ÷ 4.999A AC
0.001A ÷ 9.999A AC/DC
0.01A ÷ 29.99A AC
0.01A ÷ 39.99A AC/DC
0.01A ÷ 99.99A AC/DC
0.01A ÷ 199.99A AC
0.01A ÷ 299.99A AC
0.1A ÷ 399.9A AC/DC
0.1A ÷ 999.9A AC/DC
0.1A ÷ 1999.9A AC
0.1A ÷ 2999.9A AC
Range [W]
0.015  2.650k
0.15  10.60k
0.15  26.50k
1.5  265.0k
Range [kW]
0.000  9.999
0.00  999.99
0.0  999.9
0  9999
Range [kVAr]
0.000  9.999
0.00  999.99
0.0  999.9
0  9999
EN - 9
EASYTEST-COMBI519-COMBI521
Resolution [V]
0.1V
Resolution [mV]
0.1
Resolution [kW]
0.001
0.01
0.1
1
Resolution [kW]
0.001
0.01
0.1
1
Resolution [kVAr]
0.001
0.01
0.1
1
Quick reference guide
Accuracy
(1.0rdg + 1dgt)
Accuracy
(1.0rdg + 1dgt)
Accuracy
(2.0rdg + 2dgt)
Accuracy
(5.0%rdg + 2dgt)
Accuracy
(5.0%rdg + 2dgt)
Resolution
0.1mA AC
0.001A AC
0.001A AC/DC
0.01A AC
0.01A AC/DC
0.01A AC/DC
0.01A AC
0.01A AC
0.1A AC/DC
0.1A AC/DC
0.1A AC
0.1A AC
Accuracy
(2.0rdg + 5 dgt)
Accuracy
(2.0rdg + 5 dgt)
Accuracy
(2.0rdg + 5 dgt)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Combi519EasytestCombi521 supercombis

Tabla de contenido