The PURGE button allows the delivery of a limited volume of fluid in order to purge the extension line
prior to being connected to a patient or after changing a syringe. The purge function cannot be used if
the rate lock is enabled.
The bolus feature can be used during induction as well as during infusion.
If the volume of the bolus reaches the bolus volume limit the bolus will stop and the pump will
automatically revert to infuse at the set rate. If the volume to be infused is reached during a bolus, the
volume to be infused complete alarm will operate and the unit will revert to its previous state. Press
MUTE to stop the alarm or CANCEL to acknowledge the alarm.
Alarms are not disabled during the operation of the bolus feature. The pressure alarms are
temporarily increased to their maximum levels during the operation of the bolus feature.
A bolus cannot be administered if the rate lock is enabled or if a multidose set-up is in use.
7001FAOPT73 ISS 13
To power up the unit press the power on/off button
once.
To power off the unit press and hold down the
power ON/OFF button for 3 seconds. The display
will count down and switch the unit OFF.
1. To use the purge option stop the pump and
press the PURGE button.
2. The display will change and show two purge
softkeys. Press the two PURGE softkeys
simultaneously.
The purged volume is not added to the
volume infused.
3. When
the
completed press the QUIT softkey to exit the
menu.
1. During infusion press the BOLUS button once
to display the bolus screen.
2. Use the chevron keys to set the bolus dose
required.
3. To deliver the bolus press the two BOLUS
softkeys simultaneously. During the bolus the
unit will display the volume being infused.
4. When the desired bolus has been delivered,
release the two bolus softkeys. The bolus
volume will be added to the total volume
infused. Press QUIT to exit the bolus screen.