Getting Started
Volume To Be Infused (VTBI)
This option allows you to set a specific volume
to infuse, and rate at the end of the volume to
be infused, from stop, 1ml/h, 2ml/h and
continuous infusion at the set rate.
1. Press the "?" softkey to access the options
menu.
2. Select the SET VTBI option using the
chevron keys and press the OK softkey
indicated on the screen.
3. Enter the volume to be infused using the
chevron keys. When the volume to be infused
has been entered, press the OK softkey.
4. Select the rate at the end of the VTBI using the
rate selection chevrons. The default is STOP.
5. Press the OK softkey to exit the VTBI menu.
6. When the pump has delivered the set volume
it will alarm. Press the CANCEL softkey to
clear the alarm.
7. Press the CLEAR softkey to turn the VTBI
function off, or set a new VTBI using the
chevron keys.
Set VTBI Over Time
Set 'VTBI Over Time' must have been
activated in the Configured Options.
This option allows a VTBI and delivery time to
be specified. The rate necessary to deliver the
required volume within the required time will be
calculated and displayed. Infusion must be
stopped to select this option.
1. Press the "?" button to access the options
menu.
2. Select the SET VTBI OVER TIME option
using the chevron keys and press the OK
softkey indicated on the screen.
3. Adjust the volume to be infused using the
chevron keys, when the volume to be infused
has been entered press the OK softkey.
4. Enter the time over which the volume is to be
infused. The infusion rate will automatically
be calculated. Press the OK softkey to enter
the value.
5. Select the rate at VTBI end and press the OK
softkey. The default is STOP.
12 / 87
7001FAOPT73 ISS 13