The multidose function is enabled within the general options. See General Options in this user
manual.
7001FAOPT73 ISS 13
Getting Started
1. Press the ? button to access the options
menu.
2. Select the MULTIDOSE option using the
chevron keys
3. Press the OK softkey indicated on the
screen to confirm the selection.
4. Adjust the multidose RATE using the
chevron keys, then press the OK softkey.
5. Select the multidose VOLUME to be
infused using the chevron keys, then press
the OK softkey. If an incorrect value has
been entered, use the BACK softkey to
return to the previous stage.
6. Enter the MAINTENANCE infusion rate
using the chevron keys, then press the OK
softkey.
7. Select the FREQUENCY (hr:mins) of each
multidose using the chevron keys, then
press the OK softkey.
8. Review the multidose data on screen, then
press CONFIRM. If necessary use the
BACK softkey to return to the data. Once
confirmed, the pump will return to the main
display. The review section of the display
will
show
either
remaining or time between multidoses
remaining.
Clear Multidose
1. Stop the infusion then press the ? button.
2. Select CLEAR MULTIDOSE with the
chevron keys and press the OK softkey to
confirm.
Multidose
multidose
volume
15 / 87