Getting Started
Operating Precautions
This instrument is protected against the effects
of external interference, including high energy
radio frequency emissions, magnetic fields
and electrostatic discharge (for example, as
generated by electrosurgical and cauterising
equipment, large motors, portable radios,
cellular telephones etc.) and is designed to fail
safe if unreasonable levels of interference are
encountered.
In some circumstances the unit may be
affected by an electrostatic discharge of above
8kV. At test levels lower than these values the
unit
will
operate
normally.
In
rare
circumstances the unit may be affected by
radio frequency radiation above a level of
3V/m. If the unit is affected by this external
interference the unit will fail safe or reset, (a
call back alarm will occur after 2 minutes).
Should false alarm conditions be encountered
either, remove the source of the interference,
or regulate the infusion by another appropriate
means.
This
unit
emits
a
certain
level
of
electromagnetic radiation which is within the
levels
specified
by
IEC60601-2-24
and
IEC60601-1-2. If however the unit interacts
with other equipment measures should be
taken to minimise the effects, for instance by
repositioning or relocation.
If this instrument is dropped, subjected to
excessive
moisture,
humidity
or
high
temperature, or otherwise suspected to have
been damaged, remove it from service for
inspection by a qualified service engineer.
An explosion hazard exists if the instrument is
used
in
the
presence
of
flammable
anaesthetics. Exercise care to locate the unit
away from any such hazardous sources. An
electrical shock hazard exists if the units
casing is opened or removed. Refer all
servicing to qualified service personnel.
A comprehensive service manual containing
circuit descriptions, servicing and testing
information is available for this unit. It can be
ordered from your ALARIS Medical Systems
®
authorised
distributor
(Technical
Service
Manual Part Number 6000PB00001).
5 / 87
7001FAOPT73 ISS 13