Temperature limitation
This end up
Keep dry
Fragile handle with care
Compatibility with Radiographic Generators
GXS-700 is generally compatible with any dental X-ray unit and generator capable of
supplying the required range of exposure times and doses.
Follow the instructions of the X-ray generator to set the desired dose range.
The radiation of X-ray tubes is controlled by the settings of:
Some controls allow for the modification of all of the above, some have fixed settings for
current and voltage.
Distance of X-ray Source to Sensor
There is a correlation between the distance of the cone to the sensor and the dose received by
the Gendex GXS-700 sensor. The radiation that reaches the sensor decreases with the square
of the distance. If you double this distance, you receive only 1/4 of the radiation.
English
•
Exposure time (msec) or pulses
•
Voltage (kV or kVp)
•
Current (mA)
Specifications and Standards
7-5