Interacoustics AD528 Manual De Instrucciones página 46

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is essential to use
only the following accessories:
Item
Monitor Headset w. microphone
Bone Conductor
Audiometric Headset
Talk Back Microphone Clip-On
Free field speakers
Patient response switch
USB cable
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the cable types
and cable lengths are as specified below:
Description
Monitor Headset
w. microphone
Bone Conductor
Audiometric Headset
Talk Back Microphone Clip-On
Free field speakers
Patient response switch
USB cable
Portable and mobile RF communications equipment can affect the AD528™. Install and operate the
AD528™ according to the EMC information presented in this chapter.
The AD528™ has been tested for EMC emissions and immunity as a standalone instrument. Do not use the
AD528™ adjacent to or stacked with other electronic equipment. If adjacent or stacked use is necessary, the
user should verify normal operation in the configuration.
The use of accessories, transducers and cables other than those specified, with the exception of servicing
parts sold by Interacoustics as replacement parts for internal components, may result in increased
EMISSIONS or decreased IMMUNITY of the device.
Anyone connecting additional equipment is responsible for making sure the system complies with the IEC
60601-1-2 standard.
Guidance and manufacturer's declaration - electromagnetic emissions
The AD528™ is intended for use in the electromagnetic environment specified below. The
customer or the user of the AD528™ should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
D-0121151-B – 2018/11
AD528 - Instructions for Use - EN
Manufacturer
Sennheiser
Radioear
Radioear
G-Star
Edifier
Radioear
Sanibel
Length (m)
2.9
2.0
2.0
1.9
0.6+0.9
2.0
1.9
Compliance
Group 1
Class B
Not Applicable
Not applicable
Screened (Yes/No)
Yes
No
Yes
Yes
Yes
Yes
Yes
Electromagnetic environment - guidance
The AD528™ uses RF energy only for its internal
function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The AD528™ is suitable for use in all commercial,
industrial, business, and residential environments.
Model
PC131/PC3
B71W
DD65
TC-945
MP210
APS3
8011241
Page 42

Publicidad

Tabla de contenido
loading

Tabla de contenido