4.3.2
Carrying out manual breath sampling (optional)
If the test person cannot produce the necessary minimum volume, then sampling
can be triggered manually.
Carrying out manual breath sampling:
1.
Follow steps 1 to 3 of the automatic breath sampling.
2.
If the progress circle is displayed after starting the breath sample, briefly
press the
When the results are displayed, Manual is also shown.
4.4
Carrying out passive breath sampling (optional)
With passive sample collection, sampling is triggered manually by the user of the
instrument. Possible applications are checking the ambient air or the end-
expiratory breath of persons to check for the presence of alcohol.
Carrying out passive breath sampling:
1.
Press and hold the
appears.
2.
Hold the prepared instrument without the funnel in the air to be analysed.
3.
Briefly press the
When the results are displayed, Passive is also shown.
4.5
Removing the funnel
Hold the funnel at the gripping areas on the sides and pull it up and off the
instrument.
Dispose of the funnel in compliance with local regulations.
4.6
Menu control
Open the menu:
Press the
Navigate in the menu or save settings:
1.
Press the
2.
Confirm the selected feature or entry with the
Leave the menu:
Select the Back menu item in sub-menus.
Select the Measuring menu item in the main menu.
button.
button for approx. 1 second until the startup screen
button to trigger the measurement.
button.
or
button.
Use
button.