30

Specification

Slit Lamp SL120

Line voltage
Power consumption
Protection class
Instrument type
Enclosure protection
Environment conditions
for intended use
Storage conditions
(in original packing)
Shipment conditions
(in original packing)
Mode of operation
Light source for
projection
Lamp voltage
Fixation lamp
Magnification
Field of view
Eyepiece magnification
Slit length
Slit width
Slit rotation
Slit decentration
Swivel range of slit
prism
Angle of incidence
Filters
Free working distance
Travel of instrument
base
Vertical travel of chin
rest
Mass
100 V -10 % ... 240 V +10 %, 50 / 60 Hz
U.S. and Canada only: single-phase connection
120 V AC with plug type NEMA 5-15P
75 VA
I
B (acc. to EN 60601-1)
IP 20
Temperature:
+10 °C ...+35 °C
Humidity:
30 ... 75 % (no condensation)
Air pressure:
800 ... 1060 hPa
Temperature:
-10 °C ...+55 °C
Humidity:
10 ... 95 %
Air pressure:
700 ... 1060 hPa
Temperature:
-40 °C ...+70 °C
Humidity:
10 ... 95 %
Air pressure:
500 ... 1060 hPa
continuous operation
6 V / 20 W, halogen lamp
continuously adjustable
5 V / 1.5 W
5x, 8x, 12x, 20x, 32x (with high eyepoints 10x)
6x, 10x,16x, 25x, 40x (with high eyepoints 12.5x)
40 mm ... 6 mm (with high eyepoints 10x)
31 mm ... 5 mm (with high eyepoints 12.5x)
either 10x or 12.5x high eyepoint,
±8 D compensation of ametropia
0 to 14 mm continuously progressive
in steps at 0.3 / 3.5 / 8 / 14 mm; 1 ... 6 mm
continously
±90° continously
±4° horizontal, click-stop at 0°
180°, scale for angular difference
click-stops at -10°, 0°, +10°
0° or 0° ... 20° with tild prism head
blue, green (red-free); heat absorbing filter (fixed);
swing-in diffusor in front of the prism
65 mm
30 mm (vertical), 110 mm (X), 90 mm (Y)
58 mm
9.75 kg (basic unit), 1.25 kg (headrest)