REMOVAL OF SPREADER BAR/CRADLE SYSTEMS

FOLLOW THESE INSTRUCTIONS EXACTLY
Your presence lift has the option of using both a 6 point spreader bar and 4 point
positioning cradle. The removal of either of these systems is completed easily by the
use of a quick-release pin. This is done quickly and easily, but you must read the fol-
lowing safety instructions to ensure the spreader bar/cradle is safely locked into posi-
tion before use.
CONNECTION
To connect, raise the spreader bar/cradle up toward the boom slot.
Insert the pin sleeve through the plastic end cap and to support the spreader bar/cradle.
Once aligned in the boom slot reinsert the quick release pin fully into the sleeve until
it locks into position. YOU MUST CHECK THAT THE PIN IS SECURELY FITTED.
To do this, press your finger into the barbed end. If the pin remains stationary the lift
is safe to use.
294000.10007 Rev. C
E n g l i s h
E n g l i s h
REMOVAL
CAUTION: Be sure to support the weight of the spreader bar/cradle before removing pin.
To detach pin, depress the spring barb (using a flat blade electrical screw driver or
similar) and pull the pin out.
Support the spreader bar/cradle, and then pull the pin sleeve in the opposite direction,
removing it completely from the boom. This will release the spreader bar completely.
15
294000.10007 Rev. C