Personalized Model Transfer With The Registration Joint Holder - Ivoclar Vivadent Stratos 100 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

6.5
Personalized model transfer with the
registration joint holder
6.5.1 Return articulator to the initial set position
See 6.1.1 for description
Remove incisal indicator (1.2).
6.5.2 Mount the plane indicator
(50) instead of the incisal guide
rod (1.2).
6.5.3 Mount bite fork support
(51) to the lower frame (2.2) by
means of the instrument carrier
(40).
6.5.4 Insert registration joint
holder (52) instead of the incisal
plate (1.1) and secure it.
6.5.5 6.5.5 Mount UTS registra-
tion to the registration joint
holder and secure bite fork with
the bite fork support.
Example: Centric Tray
Example: UTS bite fork (dentulous
case)
6.5.6 Mount maxillary model to
the bite fork and cast. Close
articulator until the plane
indicator lies on the registration
joint holder.
Example: UTS bite fork
(edentulous case)
52
FH registration joint holder
(54)
The procedure is the same as with
the CP registration joint holder.
The plane indicator (50) lies on
the relevant support of the FH
registration joint holder (54).
6.5.7 Subsequently, orientate mandibular model in the usual
manner.
50
54
15

Publicidad

Tabla de contenido
loading

Tabla de contenido