Message
Comm. error with spectrometer
Comm. err. while retrieving scans
Comm. data error with spectrometer
Spectrometer calibration error
Spectrom. not receiving enough light Check sample cell for correct path length. Make
Spectrom. receiving too much light
No SD card inserted
SD card write protected
File not found on SD card
Read file error
Write file error
Parse XML file error. Corrupt file.
Interpreting XML file error. Corrupt
file.
Copying file error. Corrupt file.
Deleting file error. Corrupt file.
SD card does not belong to instru-
ment
SD card memory low, <20 MB
SD card memory too low, <2 MB
Stability test failed
Stability test outside limits
BG20 test failed
6:2
Action
Restart instrument and retry analysis. If prob-
lem persists, please contact your local service
representative.
Restart instrument and retry analysis.
Restart instrument and retry analysis. If prob-
lem persists, please contact your local service
representative.
Spectrometer needs to be re-calibrated. Please
contact your local service representative.
sure SHW is in correct position. Sample might
be too dark.
Fill sample cell. Check sample cell for correct
path length. Make sure SHW is in correct posi-
tion.
Insert SD-Card and retry.
Shift the lock switch on the SD-Card.
Reload file to SD-Card from FossManager
Updater.
Reload file to SD-Card from FossManager
Updater.
SD-Card might be full. Check write protection
on SD-Card. Repair SD-Card in FossManager
Updater.
Corrupt file. Reload file to SD-Card from Foss-
Manager Updater.
Corrupt file. Reload file to SD-Card from Foss-
Manager Updater.
Corrupt file. Reload file to SD-Card from Foss-
Manager Updater.
Corrupt file. Reload file to SD-Card from Foss-
Manager Updater.
Make sure to use correct SD-Card. Repair SD-
Card in FossManager Updater.
<20MB free space on SD-Card. Transfer data on
SD-Card to FossManager Updater.
<2MB free space on SD-Card. Transfer data on
SD-Card to FossManager Updater or oldest data
files on SD-Card will be overwritten.
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
Restart instrument and retry. If problem per-
sists, please contact your local service repre-
sentative.
User Manual 6002 5350 / Rev. 6
Infratec™ Sofia