ES
FR ENG
1. Press
key.
2. Press
key.
3. Place the calibration weight on the pan.
When the scale is stable, press
4. Remove the calibration weight from the pan and press
lly.
Put the JP3 jumper back after calibration is completed
TROUBLESHOOTING
PROBLEM
No Display
Certain figures on the
display is missing
Zero drift or span drift is
over the limit
Poor
repeatability
of
measured values.
Large fluctuation of the
displayed values
Key input is not possible
┌ - - - ┐is displayed
- L- - is displayed
- □ └ - is displayed
This scale is warranted against defects of manufacturing and materials for a period of 1 year, from the
delivery date.
During this period, GIROPÈS SL will take charge of repairing the scale.
This warranty does not cover defects or damaged caused by misuse, overloading or improper installation
contrary to the recommendations described in this manual.
This warranty does not cover shipping costs for the reparation of the balance.
User's guide
All manuals and user guides at all-guides.com
pkey to cycle through a list of permissible weights.
POSSIBLE CAUSES
_ AC adaptor is broken
_ Main board is broken
_ Main board is broken
It is directly exposed to sunlight or loca-
ted close to extreme temperature such
as air-conditioner or heater
_ Wind and vibration
_ Load cell is touch something
_ Dirt or dust on the load cell
_ Main board is damaged
Keypad was not connected well.
Main board is damaged.
Overload
_ Weighing pan is not in place
_ Weighing range is below zero
Zeroing outside the zero setting range
key. The instrument will restart automatica-
SOLUTIONS
_ Replace the AC adapter
_ Replace the main board
_ Replace the main board
_ Check if the LCD is soldered well
Move the instrument to a mild tempera-
ture installation site.
_ Change the installation site.
_ Check the load cell and the wire lead
assembly.
_ Clean the dirt or dust on the load cell
_ Replace the main board
Check the keypad connection.
Replace the main board
Clear the pan or reduce the weight.
_ Ensure the weighing pan is correctly
installed and the accessories are not
touching the weighing pan or load cell
_ Reset the scale to zero
Ensure that zeroing is performed in the
permissible range. (20% of capacity)
4. WARRANTY
MPI
17