Calibration Procedure (Cal2) - Baxtran MPI-6 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ES
FR ENG
tA
3.
- TARE FUNCTION
DISPLAY
tA
0
tA
1
Pr
4.
- RS232 OUTPUT
DISPLAY
Pr
LP-50
Pr
PC
br
5.
- BAUD RATE
DISPLAY
Br 2400
Br 4800
Br 9600
CALIBRATION PROCEDURE (CAL 2)
Notice: This section can only be operated by engineers.
Please have the jumper JP3 switch OFF before you start the calibration process.
In Calibration procedure, press
1. SEL MAXIMUM CAPACITY
DISPLAY
LoAd 06
LoAd 15
LoAd 30
2. DECIMAL POINTS
DISPLAY
Dp 0
Dp 0.0
Dp 0.00
3. SPAN CALIBRATION
DISPLAY
SPAn
User's guide
All manuals and user guides at all-guides.com
EXPLANATION
One time taring
Multiple taring is allowed
EXPLANATION
Parity: 8-N-1, LP-50 printer format (Baud rate must be 9600).
Parity: 8-N-1, connecting to PC.
EXPLANATION
Set the baud rate to 2400
Set the baud rate to 4800
Set the baud rate to 9600
to toggle among the options, and press
MAXIMUM CAPACITY
6 kg
15 kg
30 kg
DESCRIPCIONES
Disable decimal points.
Set the price display to 1 decimal point.
Set the price display to 2 decimal points.
EXPLICACIÓN
The Weight column and the Unit Price column display the Span Value.
MPI
to save the setting.