Fault
The controller cannot be switched on
(no beep, no orange or green light
signal).
After switching the controller on, no
aerosol is generated, and/or the
controller switches off again after
a few seconds.
The controller switches off even
though there is still medication left in
the medication reservoir.
Info:
Note that about 1 ml of medica-
tion cannot be nebulised and
remains unused in the medica-
tion reservoir of the
®
eFlow
rapid nebuliser handset;
this residue must be discarded.
This is intentional and does not
represent a malfunction.
With medication-specific nebu-
liser handsets (Tolero
®
and Altera
) there is no signifi-
cant volume of residue left in
the medication reservoir.
The controller does not switch off
automatically even when no more
aerosol is being produced and there
is no more than about 1 ml of residue
left in the medication reservoir of the
®
eFlow
rapid nebuliser handset.
®
eFlow
rapid - 2021-04
Possible cause/Remedy
- To switch on, press the ON/OFF button
for approx. 2 seconds.
- Is power plug plugged into mains and
the device (for mains operation)?
- Are batteries inserted correctly?
- Check the charge level of the batteries.
- Has the medication reservoir been filled
with medication?
- Check the connection between the
controller and the nebuliser handset.
If there is still more than 1 ml in the
medication reservoir of the eFlow
nebuliser handset:
- Check the charge level of the batteries.
- Was the nebuliser handset held
horizontally throughout the inhalation
session? Hold the nebuliser handset
horizontally.
- The controller switches off
automatically after 20 minutes.
Carry out inhalation to the end by press-
ing the ON/OFF button again, so that
you can inhale the prescribed dose of
®
®
medication.
, Zirela
If the maximum operating time for each
inhalation session is exceeded
repeatedly, the aerosol head must be
replaced.
The treatment session has been com-
pleted successfully and the controller can
be switched off by pressing the ON/OFF
button.