TROUBLESHOOTING

6
You can correct malfunctions that occur during daily operation yourself. The following
tables will help you to locate and correct the causes of malfunction.
Display indicator
(flashes alternately)
54
Possible cause
Cable connection to
the controller or to
the nebuliser
handset interrupted.
No medication
added.
Inhalation
interrupted.
The maximum
operating time of
20 minutes for an
inhalation session
has been exceeded.
Battery charge level
50%.
Battery flat.
Solution
Check cable
connections.
Add medication.
Resume inhalation
by restarting the
device (press ON/
OFF button).
Resume inhalation
by restarting the
device (press
ON/OFF button).
Have new batteries
ready.
Insert new batteries
or operate with the
power adapter
connected.
®