• Take the parts out of the disinfector
immediately after disinfection is com-
plete. Place them on a dry, clean sur-
face and allow them to air dry
completely.
CAUTION:
Check the nebuliser handset parts after
every disinfection. Replace all broken,
deformed and heavily discoloured parts.
Possible option: Boiling in distilled
water
NOTE:
Plastic will melt if it comes into contact with
the hot base of the saucepan. Therefore,
make sure that there is sufficient water in
the saucepan. This will prevent the parts
from being damaged.
• Put all parts in boiling water for at least
5 minutes. Use a clean saucepan and
distilled water for this.
• Take the parts out of the saucepan.
• Shake all parts to remove excess water
more quickly.
• Place the individual parts on a dry, clean
surface and allow them to air dry
completely.
CAUTION:
Moisture promotes the growth of bacte-
ria. Please ensure adequate drying after
every disinfection.
CAUTION:
Check the nebuliser handset parts after
every disinfection. Replace all broken,
deformed and heavily discoloured parts.
52
Possible option: Chemical disinfection
• Place all nebuliser handset parts and
the aerosol head in the chemical disin-
fectant solution. Pay attention to the
information regarding suitable chemical
disinfectants and the instructions for use
of the disinfectant (see "Possible alter-
native: Chemical disinfection",
page 53).
• Place all individual parts on a dry, clean
surface and allow them to air dry
completely.
CAUTION:
Moisture promotes the growth of bacte-
ria. Please ensure adequate drying after
every disinfection.
CAUTION:
Check the nebuliser handset parts after
every disinfection. Replace all broken,
deformed and heavily discoloured parts.
Hygienic reprocessing in the
hospital and doctor's
practice
Carry out cleaning with disinfection
immediately after use. Ideally, a mechani-
cal procedure (instrument dishwasher)
should be used.
Prepare for cleaning with disinfection (see
"CLEANING AND DISINFECTION",
page 50).

Cleaning with thermal disinfection

• Place the nebuliser handset compo-
nents in the instrument dishwasher.
• Select the 93 °C programme (treatment
time: 10 minutes).
The effectiveness of this procedure has
been validated using a G7736 disinfector
made by Miele and using neodisher MA
(Chemische Fabrik Dr. Weigert, Ham-
burg) as a cleaning agent in conjunction
with deionised water as a neutraliser.
®
eFlow
rapid - 2021-04