C C o o v v e e r r r r e e p p l l a a c c e e m m e e n n t t
C C A A U U T T I I O O N N - - Always take caution when you replace the support surface cover. The internal fire barrier contains fiberglass
fibers. Dust particles from the fibers may cause skin irritation.
T T o o o o l l s s r r e e q q u u i i r r e e d d : :
None
P P r r o o c c e e d d u u r r e e : :
1. Raise the bed height to the full up position.
2. Lower the Fowler and Gatch sections to the full down positions.
3. Unzip the cover. Start at the foot end patient right corner of the support surface and stop at the head end patient right
corner.
4. Fold the top of the cover to the patient's right side, and then remove the foam crib assembly from the bed.
5. Remove and discard the cover.
6. Place the replacement cover, unzipped and open, on the bed with the black bottom cover on the litter and the top cover
folded over the patient's right side of the bed.
7. Place the foam crib assembly on top of the bottom part of the cover to align the foam crib with the cover.
8. Fold the top cover over the top of the foam crib assembly to align that the top cover with the foam crib assembly.
9. Zip the cover to close. Start at the head end patient right corner and stop at the foot end patient right corner.
10. Verify proper operation before you return the product to service.
EN
12
2815-009-005 Rev AC.0