6 Equipment description
Presoclave III Plus autoclaves belong to a series of saturated vapour autoclaves
that are distinguished for the following features:
1. Arm / joist locking system.
2. Temperature and time electronic control circuit, and process registration. 10 user-
definable programs.
3. Safety locking system that only allows lid opening with pressure 0 bar and a
temperature under 95ºC.
4. Maximum pressure limiter that disconnects the heating element in case of over-
pressure, set at 2.5bar.
5. Limiter thermostat to protect the heating element, set at 160ºC.
6.1 Equipment
-
Safety system:
Safety over pressure valve.
Safety pressure switch.
Safety thermostat.
Lid closed detector.
Protective thermal lid cover.
-
Chamber and lid made in stainless steel.
-
External case made in stainless steel.
-
Manual emptying valve.
-
Waste water tank connection.
6.2 Features
-
Sterilization temperature from 115°C to 134°C.
-
Temperature resolution 1ºC.
-
Sterilization time from 3 minutes to 180 minutes (3h)
-
Time resolution 1 minute.
-
Suitable to work with solids and liquids atmospheric purge from 115°C to 134°C.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España