6. Equipment description

6.1
Solid load and atmosferic air removal SOL1:
Solid with two forzed vacuum pulses SOL2:
Solid with three forzed vacuum pulses SOL3: 115-134ºC
Liquids load, atmosferic air removal LIQ:
Sterilizing time range:
Drying time range:
6.2
Autoclave is delivered with 10 presset cycles, for sterilization and two test cycles.
Nr Name
0
Not warpped load
1
Wrapped load
2
Resistant
3
Plastic not wrapped
4
PLastic wrapped
5
Hollow devices
6
Desinfección
7
Liquid (media)
8
Metal not wrapped
9
Metal not wrapped
6.3
Test and validation cycles from EN 13060.
Test de Bowie Dick
Vacuum Test
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Cycle types and temperature range
Pre-set cycles
Temperature

Test cycles

Heart probe
105-134ºC
105-130ºC
115-134ºC
-
-
105-134ºC
105-130ºC
From 3 to 59 minutes
From 0 to 99 minutes
Dry
Time
ºC
(minutes)
134
4
10 2x vacuum cycles
134
12
30 3x vacuum cycles
134
20
20 3x vacuum cycles
121
15
15 2x vacuum cycles
121
30
30 3x vacuum cycles
126
12
25 3x vacuum cycles
105
25
15
121
30
-
121
15
30
134
4
30
134ºC
Lower vacuum level
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Page: 22
Purge
Atmosferic
Atmosferic
Atmosferic
Atmosferic
3' 30''
5+ 10'