English
7. MAINTENANCE
7.1 General
1. Check main power cord and plug if there is abrasions or excessive wear.
2. Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are
stubbed together correctly.
3. Check airflow from the quick connector. The airflow should alternate between each connector
every half-cycle time if it's in alternating mode.
4. Check the air hoses if there is kink or breaks. For replacement, please contact our local dealers.
7.2 Fuse Replacement
7.3 Air Filter Replacement
8. EXPECTED SERVICE LIFE:
The products are intended to offer safe and reliable operation when use or installed according to the
instructions provided by Apex Medical. Apex Medical recommends that the system be inspected and
serviced by authorized technicians if there are any signs of wear or concerns with device function and
indication on products. Otherwise, service and inspection of the devices generally should not be
required.
1. Disconnect the plug form mains power
when a blown fuse is suspected.
2. Remove the cover of the fuse holder by
means of a small screwdriver.
3. Insert a new fuse of the correct rating in,
and replace the cover of the fuse holder
back. The fuse should be rated as
T1AL/250V type and VDE approved.
1. Loosen the screws from the air filter plate
located at the back of pump.
2. The filter is reusable and can be washed
gently with a mild detergent and water. Dry
the filter before use.
3. Check and replace air filter regularly if
environment is dirty.
12