Apex Pro-CARE 4 Manual De Uso página 18

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 20
English
Guidance and Manufacturer's Declaration - Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a) The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795
MHz;13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
b) The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80
MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable communications equipment could
cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is
used in calculating the recommended separation distance for transmitters in these frequency ranges.
c) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
IEC60601 test level
Compliance
3Vrms150 kHz to 80
3 Vrms
MHz outside ISM
a
bands
3 V/m 80 MHz to 2.5
3 V/m
GHz
16
Electromagnetic Environment-Guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of this device, including cables,
than there commended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
=
d
1
2 .
P
150kHz to 80MHz
=
d
1
2 .
P
150kHz to 80MHz
=
d
2
3 .
P
80 MHz to 2.5G MHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
b
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
c
survey
, should be less than the
compliance level in each frequency
d
range
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

9p-052500

Tabla de contenido