Error Codes; Accessories; Product Contact Parts - IKA Vacstar digital 112017 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47
If an error occurs, this is shown in an error on the display (B), e.g. Error 4.
Proceed as follows:
Switch the device off on the device switch,
Take corrective measure,
Switch device back on.
Error
Cause
Error 3
Overheat
Error 4
Motor blocked/overload
Error 8
Hall sensor defective
Error 9
BLP self-test/EEPROM error
No display
Pump diaphragm defective
No display
Diaphragm valve defective
No display
Liquid in the pump chambers
In the event that the fault cannot be eliminated using the measures described or another error code is displayed:
Please contact the Service Department,
Send the device and a brief description of the fault.
Vacuum controller IKA VC 10
Vacuum control valve IKA VCV 1 and VCV 2
Check valve IKA VC 10.300
Vacuum hose IKA VH.SI.8
PC 1.1 cable (RS 232)
Name
Connection spigot
Connection piece
Distributor
Head piece
Diaphragm valve
Diaphragm
Clamp

Error codes

Reason
The pump does not run
The pump does not run
The pump does not run
The pump does not run
Minimum pressure and suction
rate are not achieved
Minimum pressure and suction
rate are not achieved
Minimum pressure and suction
rate are not achieved

Accessories

For further accessories see www.ika.de

Product contact parts

Material
PPS
PPS
PPS
PPS
FFPM
NBR/PTFE
PPS
Solution
Switch off the pump, disconnect it from the mains,
let it cool down.
Switch off the pump, disconnect it from the mains
Let the pump cool down.
Ventilate the vacuum system to normal pressure.
Service
Service
Service
Request a maintenance kit
Service
Request a maintenance kit
Let the pump run dry (no-load operation)
Maintenance kit
Vacuum safety emission condenser VSE 1
Woulff bottle/water valve set VSW1
29

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

20000016633

Tabla de contenido