How This Ear Thermometer Measures Temperature; Control Displays And Symbols - Microlife IR 100 Manual Del Usuario

Ocultar thumbs Ver también para IR 100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 14
• Never immerse this instrument in water or other
liquids. For cleaning please follow the instruc-
tions in the «Cleaning and Disinfecting» section.
• Do not use the instrument if you think it is damaged or
notice anything unusual.
• Never open the instrument.
• Earwax in the ear canal may cause a lower temperature
reading. Therefore it is important to ensure the subject's
ear canal is clean.
• This instrument comprises sensitive components and
must be treated with caution. Observe the storage and
operating conditions described in the «Technical Specifi-
cations» section.
• Protect it from:
− extreme temperatures
− impact and dropping
− contamination and dust
− direct sunlight
− heat and cold
• If the instrument is not going to be used for a prolonged
period the battery should be removed.
• Ensure that children do not use the instrument unsuper-
vised; some parts are small enough to be swallowed.
4
WARNING: Use of this instrument is not intended
as a substitute for consultation with your physician.
This instrument is NOT waterproof so do not
immerse in liquids.
3. How this Ear Thermometer measures
Temperature
This thermometer measures infrared energy radiated from
the eardrum and the surrounding tissue. This energy is
collected through the lens and converted to a temperature
value. The measured reading obtained directly from the
eardrum (Tympanic Membrane) ensures the most accurate
ear temperature.
Measurements taken from the surrounding tissue of the ear
canal generate lower readings and may result in misdiag-
nosis of fever.
To avoid an inaccurate measurement:
1. Switch on the thermometer by pressing the ON/OFF
button 4.
2. After one beep is heard (and the temperature scale icon
is flashing), straighten the ear canal by gently pulling the
middle of the ear back and up.
3. Place the probe 1 firmly into the ear canal, press the
START button 2 and keep the probe in the ear until the
thermometer beeps to signal the completion of the
measurement.

4. Control Displays and Symbols

• All segments displayed 6: Press the ON/OFF button
4 to turn on the unit; all segments will be shown for
2 seconds.
• Memory 7: The last reading will automatically be
shown on the display for 2 seconds.
• Ready for measurement 8: When the unit is ready for
measurement, the «°C» or «°F» icon will keep flashing.
• Measurement complete 9: The reading will be
shown on the display 3 with the «°C» or «°F» icon; the
unit is ready for the next measurement, when the «°C»
or «°F» icon is flashing again.
• Out-of-ear temperature indication AT: A crossed-
ear-icon will appear on the display 3 if the reading falls
outside the range 32.0 ∼ 42.2 °C (89.6 ∼ 108.0 °F).
• Low battery indication AK: When the unit is turned on,
the battery icon will keep flashing to remind the user to
replace the battery.

Publicidad

Tabla de contenido
loading

Tabla de contenido