Descargar Imprimir esta página

Megger IDEAL 61-791 Manual Del Usuario página 5

Probadores manuales de aislamiento y continuidad

Publicidad

Idiomas disponibles

Idiomas disponibles

voltmeter on all switch positions except 'OFF' and
will sound on all switch positions except 'OFF' and
inhibited except for Insulation tests, which will remain available until the voltage
exceeds 55 V.
Default Voltmeter Ranges: 25V - 450V a.c. 50/60 Hz ± 2% ±3 digits
Automatic Discharge: Capacitive circuits are automatically discharged
when the test button is released following an insulation test.
Power Supply:
Battery life:
Battery cells should not be left in an instrument which may remain unused for
extended periods of time.
Low Battery Indicator: The low battery indicator symbol
Battery Replacement:
leads are connected. To remove the rear cover, release the screw at the
bottom of the cover and lift the cover upwards.To avoid the possibility of
shock, do not press the test button or touch the fuse when changing
batteries.
Fuse:
To check this fuse, select MΩ, open circuit the test leads and press the test
button until a reading is obtained. Display of the fuse symbol
code indicates a ruptured fuse. Located behind the rear cover, this fuse can
be replaced by the user. The rear cover must not be opened if test leads are
connected. The replacement fuse must be of the correct type and rating. To
avoid the possibility of shock, disconnect the battery before touching the fuse.
Safety:
The instruments meet the requirements for double insulation to IEC
1010-1 (1995) EN 61010-1 (1995) to Category III*, 300 Volts phase to earth,
440 Volts phase to phase, without the need for separately fused test leads.
*Relates to transient overvoltage likely to be found in fixed installation wiring.
E.M.C.
Weight:
Dimensions:
Cleaning:
Supplied
Synthetic zip-up pouch
Test leads, prods & crocodile clips
450V - 600V a.c. 50/60 Hz ± 3%
d.c. - unspecified
6 x 1,5V cells IEC LR6 type only.
Typically 3000 x 5 second operations (worst
case, with
not selected).
when the battery cells are exhausted.
The rear cover must not be opened if the test
500 mA (F) H.B.C.10 kA min (32mm x 6mm)
Meets EN 50081-1 and EN 50082-1 (1992).
530g (including batteries)
195mm x 98mm x 40mm.
Wipe disconnected instrument with a clean cloth
dampened with soapy water or Isopropyl Alcohol
(IPA).
ACCESSORIES
. In addition, the beeper
. All selected tests will be
will appear
or an error
Part Number
C90
TL - 791

Publicidad

loading