Endress+Hauser Micropilot NMR81 Manual De Instrucciones página 96

Ocultar thumbs Ver también para Micropilot NMR81:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Diagnóstico y localización y resolución de fallos
96
Número de
Texto corto
diagnóstico
333
System recovery required
334
System recovery failure
381
Displacer distance invalid
382
Sensor communication
Diagnóstico de la configuración
400
AIO simulation output
401
DIO simulation output
403
Calibration AIO
404
Calibration AIP
405
COMM timeout DIO 1 ... 8
406
IOM offline
407
COMM timeout AIO 1 ... 2
408
Invalid range AIO 1 ... 2
409
RTD temp out of range
1 ... 2
410
Data transfer
411
Hart device 1 ... 15 has
malfunction
412
Processing download
413
NMT 1 ... 15: element is
open or short
415
Hart device 1 ... 15 offline
434
Real time clock defective
436
Date/Time incorrect
437
Configuration
incompatible
Remedio
HW change detected
System configuration
recovery required
Go to menu on device and
perform recovery
HW changed, system
recovery failure. Return to
factory
1. Calibrate sensor
2. Restart device
3. Replace sensor electronics
1. Check connection of
sensor electronics
2. Restart device
3. Replace sensor electronics
Deactivate simulation AIO
output
Deactivate simulation DIO
output
1. Restart device
2. Change I/O module
1. Restart device
2. Change I/O module
1. Check wiring
2. Change I/O module
1. Check wiring
2. Change I/O module
1. Check wiring
2. Change I/O module
1. Check device
configuration.
2. Check wiring.
1. Check electronic modules
2. Change I/O or main
electronic module
1. Check connection
2. Retry data transfer
1. Check HART device
2. Change HART device
Download active, please wait
1. Check NMT wiring
connection
2. Replace NMT
1. Check HART device
2. Change HART device
Replace main electronics
Check date and time settings. M
1. Restart device
2. Contact service
Micropilot NMR81
Señal de
Comportamiento
estado
de diagnóstico
[Ex-
[Ex-fábrica]
fábrica]
F
Alarm
F
Alarm
F
Alarm
F
Alarm
C
Warning
C
Warning
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
C
Warning
C
Warning
F
Alarm
1)
F
Alarm
C
Warning
C
Warning
C
Warning
C
Warning
Warning
F
Alarm
Endress+Hauser