Interconnections; P3 - Vent; P4 - External Fan; J1 - In-Out - Varian TV 1001 Navigator Manual De Instrucciones

Tabla de contenido

Publicidad

INTERCONNECTIONS

The following figure shows the TV 1001 intercon-
nections.

P3 - Vent

This is a dedicated 24 Vdc connector to control the
optional vent valve.
P4 – External fan
This is a dedicated 24 Vdc connector to supply the
optional external fan.

J1 - IN-OUT

This connector carries all the input and output sig-
nals to remote control the TV 1001 Navigator.
It is a 15-pins D type connector; the available sig-
nals are detailed in the table, the following para-
graphs describe the signal characteristics and use.
PIN N.
1
START/STOP (+)
2
START/STOP (-)
3
INTERLOCK (+)
4
INTERLOCK (-)
5
SPEED SETTING (+)
6
SPEED SETTING (-)
7
SOFT START (+)
8
SOFT START (-)
9
+24 Vdc
10
SPARE
11
PROGRAMMABLE SET POINT
12
SPARE
13
FAULT OUTPUT
14
PROGRAMMABLE ANALOG
SIGNAL (+)
15
GROUND
PROGRAMMABLE ANALOG
SIGNAL (-)
Signal description
START/STOP: input signal to start or stop the
pump. With the supplied cover connector the
START/STOP (+) signal is connected to the +24
Vdc pin and the START/STOP (-) signal to the
GROUND pin: in this condition the pump automati-
cally starts as soon as the controller recognises
the input supply ("Plug & Pump").
INTERLOCK: input signal to control the pump
rotation. With the supplied cover connector the IN-
TERLOCK (+) signal is connected to the +24 Vdc
pin and the INTERLOCK (-)signal to the GROUND
pin.
75
TECHNICAL INFORMATION
SIGNAL NAME
OUTPUT
87-900-945-01(G)
INPUT/
IN
IN
IN
IN
IN
IN
IN
IN
OUT
OUT
OUT
OUT
OUT

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Tv 551Tv 701969-8822

Tabla de contenido