Setting Up The Scale; Aligning The Scale - Seca 634 Manual De Instrucciones Y Garantías

Ocultar thumbs Ver también para 634:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 203

5.5 Setting up the scale

Aligning the scale

2 .
Spirit level
Air bubble
74 •
2. Insert the connector of the power pack into the
connecting socket of the scale.
3. Plug the power pack into a power supply socket.
1. Place the scale on firm, level surface.
2. Undo the knurled wheels.
3. Level the device by turning the foot screws.
4 .
3 .
The air bubble in the spirit level must be located in
the exact center of the circle.
4. Tighten the knurled wheels in the direction of the
arrow.
The foot screws are secured against being
adjusted.
ATTENTION!
Incorrect measurement due to force shunts
If the scale and housing is in contact with some-
thing, e.g. a towel, weight will not be measured
correctly.
– Set the scale up so that only its foot screws
are in contact with the floor.

Publicidad