4.2 Calibration of the sensor (zero reset)
1.
Wait until the sensor has been stable for at least 30 seconds.
2.
Preferably place the sensor outside in a CO-free zone, remove the
CAL screw protective cap and reset the display to 0 +- 1 ppm using
the screwdriver provided.
If you are unsure of the quality of ambient air, you can apply a
mixture of Oxygen / Nitrogen to the sensor in order to reset to zero.

5. GENERAL CHARACTERISTICS

Type of sensor:
Calibration of the sensor:calibrated by the manufacturer with a
Measuring range :
Accuracy:
Response time:
Long-term drift:
Climatic environment:
- Use:
- Storage:
Power supply:
Dimensions / Weight: 173 x 60.5 x 38 mm / 190 g with battery
Electromagnetic compatibility:
Impermeability:
Only use specified spare parts for maintenance purposes. The
manufacturer cannot accept any responsibility for accidents occurring
following repairs carried out outside its after-sales department or
approved maintenance network.
6.1

Cleaning and servicing

6.1.1

Changing the battery

Set the switch to OFF
Remove the cover at the back of the instrument
Replace the old battery with a 9V battery (6LR61 or 6LF22 type)
electrochemical, specifically for CO.
source at 205 ppm.
Nominal: from 0 to 1000 ppm (2000 ppm for a
maximum period of 5 minutes).
+- 5% of the reading +- 5 ppm (from 0 to 1000
ppm).
< 70 seconds to 90% of the measurement
<5% per year, depending on the frequency of use
0 to 50°C (32 to 122 °F) and from 15 to 90%
humidity, without condensation.
-20 to +60 °C (-4 to +140°F) and from 0 to 80
RH, without the battery.
9V battery (6LR61 or 6F22 type)
Emission as per EN 50081- Ed. 1992
Immunity as per EN 50082- Ed. 1992
IP44 as per IEC 60259 ( Ed. 89), the instrument
must be in its protective case.

6. MAINTENANCE