K. SPECIFICATIONS: IEC TABLES
The Following Tables Are Provided In Accordance With IEC 60601-1-2:
between portable and mobile RF communications equipment and the
The Vios
®
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the Vios
mobile RF communications equipment (transmitters) and the Vios
a minimum as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection f rom
structures, objects and people.
The Vios
is intended for use in the electromagnetic environment specified below. The customer or the user of the Vios
®
should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
EN-9
Questions? 1.800.FAST.NEB (327.8632) | PARI.com/VIOS
Recommended separation distances
can help prevent electromagnetic interference by maintaining a minimum distance between portable and
®
150 kHz to 80 MHz
W
d =
0,01
0,1
1
10
100
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A / m
can help prevent electromagnetic interference by maintaining
®
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
3,5
P
d =
V
1
.12
.12
.37
.37
1.17
1.17
3.69
3.69
11.67
11.67
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A / m
Vios
®
m
800 MHz to 2,5 GHz
3,5
P
d =
E
E
1
.23
.74
2.33
7.38
23.33
Electromagnetic environment
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Vios
requires continued
®
operation during power mains
interruptions, it is recommended
that the Vios
be powered from
®
an uninterruptible power supply
or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.