SHAKING INCUBATOR
SI500
Instructions for use
1. Introduction
Thank you for purchasing this piece of Stuart
equipment. To get the best performance from
the SI500 please read these instructions carefully
before use. Before discarding the packaging
check that all parts are present and correct.
For your own safety and that of others
please read and understand the safety
advice given below before using the
equipment.
2. Safety advice before use
If the equipment is not used in the manner
described in this manual and with
accessories other than those recommended by
Stuart, the protection provided might be
impaired.
This equipment is designed to operate under the
following conditions:
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5ºC to +40ºC
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply fluctuation not exceeding 10%
❖ Over-voltage category II IEC60364-4-443
❖ Pollution degree 2
❖ Use with a minimum distance all around of
200mm from walls or other items
❖ Not to be used to heat or mix materials which
could lead to liberation of gases, explosions,
implosions
❖ Not to be (used or) placed in areas known to
be hazardous
Do not attempt to lift or carry the unit alone.
Never move or carry the unit when in use or
connected to the mains electricity supply.
Do not use in a hazardous atmosphere or with
hazardous materials.
Ensure that the load is balanced.
Do not attempt to stop movement by hand.
In the case of mains interruption, the unit will
not restart on restoration of the electricity supply.
In the case of mechanical interruption, (e.g.
motor stall), the unit will continue operating on
removal of interruption.
Mechanical energy can lead to breakage of glass
vessels. Use with care.
3. General description
The Stuart SI500 orbital incubator is ideal for
mixing and aerating various biological samples
under accurate temperature controlled
conditions. The rotating platform has a digital
speed setting providing a smooth, uniform
circular motion with an orbit of 16mm. The
speed range is 30 to 300rpm. There is also an
adjustable digital countdown timer that
automatically stops after the set time (1 second
steps to 1h, 1 minute steps to 24h or 1h steps to
9 days). The shaking platform has a drawer
mechanism for easy access to the incubation
vessels. The maximum load is 10kg. The
temperature is controlled by a simple "press to
set" system in 0.1ºC intervals from ambient
+5ºC to 60ºC.
The SI500 comes fitted with a fully adjustable
cradle system to accommodate a wide range of
flasks and bottles (maximum size 1l flask).
Optional accessories include racks to
accommodate 1.5ml, 15ml and 50ml tubes.
4. Preparation for use
4.1 Electrical installation
THIS INSTRUMENT MUST BE EARTHED
Before connecting the instrument please read
and understand these instructions and ensure
that the line supply corresponds to that shown
on the rating plate. The instrument is designed
for use on 230V 50Hz.The power consumption
of the unit is 300W. The instrument is fitted with
an IEC socket at the rear of the instrument for
connection of the mains lead.
1