5.2 Setting the timer

In order to use the timer, stop the rotation and
press the mode button so that the red dot in the
display appears below "time". The display will
show three horizontal lines or the last stored
time. Select the desired countdown time using
the control knob. The time will be displayed first
in minutes and seconds, next in hours and
minutes then in days and hours up to a
maximum of 9 days.
When the display shows the correct time, press
the start/stop button to begin operating. During
operation, motion can be paused at any time by
pressing the start/stop button. If the button is
pressed again, motion will start again and the
timer will continue to count down. The timer can
be adjusted without halting the unit. When the
timer reaches zero, the unit will automatically
stop and an alert will sound. If the door is
opened while the unit is counting down, a
warning alert will sound and the motion and
countdown will stop; the word "door" will also
be momentarily displayed on the screen. Close
the door to continue and the countdown will
resume from the point at which it was stopped.
N.B to disable the timer and revert to
continuous rotation, select - - - in the
display during timer mode.

5.3 Setting the temperature

Using the top control panel, switch the
temperature control unit ON by pressing the
control knob. Holding down the "press to set"
button, turn the knob until the correct
temperature is displayed. Release the "press to
set button". The incubator will then begin to
heat to the set temperature. The actual
temperature inside the unit will be displayed on
the screen. The minimum set temperature is
ambient +5ºC and the maximum, 60ºC.
Press to set
temperature °C
press to set
Control knob
on/off
Red digital display
5.4 USB Connectivity
Your SI500 has capability to connect to a PC via
USB for long term tracking of the incubator
temperature, whilst left overnight for instance.
Remove the waterproof cap from the USB port
on the SI500 and connect the included cable,
connect the other end of the cable to a free USB
port on your PC. The latest version of our SI500
software is available for download from
www.barloworld-scientific.com, once
downloaded select run the program and follow
the on screen commands.
6. Maintenance, servicing
and repair
WARNING: Ensure the unit is disconnected
from the mains electricity supply before
attempting maintenance or servicing.
This equipment does not require routine
servicing. The only maintenance required is to
clean internal and external surfaces with a damp
cloth and mild detergent. Spillages or splashes
should be cleaned up straightaway after isolating
the unit from the mains electricity supply.
6.1 Repairs and support
Any repairs or replacement of parts MUST be
undertaken by suitably qualified personnel.
Only spare parts supplied or specified by
Stuart or its agent should be used. Fitting of
non-approved parts may affect the
performance of the safety features designed
into the instrument.
For a comprehensive list of parts required by
service engineers conducting internal repairs,
or to receive a service manual, please
contact the Service Department;
service@barloworld-scientific.com
tel: +44 (0)1785 810475
fax: +44 (0)1785 810471 quoting both the
model and serial number. For any other technical
enquiries, please contact the Technical Support
Department at;
stuarthelp@barloworld-scientific.com
tel: +44 (0)1785 810433.