Certificate: 80000336
Project: 80000336
CONDITIONS OF ACCEPTABILITY
(1) Equipment is intended to be used with power supply FSP Group Inc, Model FSP008-P01N (Sartorius
model YEPS01-15V0W). The equipment can also be used with any other power supply approved for
Canada and USA under the 60950-1 or 61010-1 or equivalent standards and which provides Reinforced
insulation between mains and secondary circuits. It must be rated for 3000m and up to 40°C and must
provide an output of 15Vdc LPS (must comply with clauses 6.3 and 6.5 of 61010-1) with a minimum of
0.3A.
(2) External Direct plug-in power supply was considered as the disconnect device.
(3) For use in dry locations only.
(4) No user replaceable battery. Internal lithium coin cell can only be replaced by Service personnel.
(5) The display YSD01 is an accessory (no CSA label) for the scale and can be used as a remote (second)
display, which is connected to the scale via a USB cable.
APPLICABLE REQUIREMENTS
CAN/CSA-C22.2 No. 61010-1-12
UL Std. No. 61010-1 (3
DQD 507 Rev. 2019-04-30
-
Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use, Part 1: General Requirements
rd
Edition)
- Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use - Part 1: General Requirements