WIKA FSD-3 Manual De Instrucciones página 28

Detector de caudal
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

9. Specifications
Specifications
Voltage supply
Power supply
Current consumption
EN
Total current consumption
Accuracy specifications
Non-repeatability
Accuracy at reference conditions
Temperature error at -20 ... +85 °C
(-4 ... +185 °F)
Reference conditions
Ambient temperature
Atmospheric pressure
Humidity
Medium
Power supply
Mounting position
Load
28
DC 15 ... 35 V
Switching output with analogue signal: 175 mA
Switching output without analogue signal: 150 mA
max. 650 mA including switching current
Flow (5 ... 100 cm/s): ±2 cm/s
Temperature: 0.5 K
Flow (5 ... 100 cm/s): ≤ ±5 % of end value of measuring range
Flow (>100 ... 175 cm/s): ≤ ±10 % of end value of measuring range
Temperature: ≤ ±1,5 K
Including non-linearity, hysteresis, zero offset and end value deviation (corresponds to
measured error per IEC 61298-2).
The accuracy of the flow switch is dependent on the thermal conductivity and the contamination
of the probe.
≤ ±0.4 cm/s per K
15 ... 25 °C (59 ... 77 °F)
860 ... 1,060 mbar (12.47 ... 15.38 psi)
45 ... 75 % r. h.
Water
DC 24 V
Process connection M18 x 1.5 downwards
Inner diameter of pipe 26 mm
Upstream/Downstream pipe: 1 m / 0.5 m
Marking towards the upstream side twist of ±5°
100 Ω
WIKA operating instructions flow switch, model FSD-3

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido