HealthSmart 04–695–001 Manual De Instrucciones página 26

Ocultar thumbs Ver también para 04–695–001:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Table 3 - Continued
For all ME EQUIPMENT and ME SYSTEMS
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
in the location in which the 04-695-001 is used exceeds the applicable RF
compliance level above, the 04-695-001 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the 04-695-001.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
26

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido